Users Manual Part 4
IBA | 35-12 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Preparing the Equipment for Irradiation
|
Nozzle devices prepared
For accessories:
Accessories list confirmed by user: the checkmark is present when you
have put a checkmark in the checkbox at the bottom of the tab, to confirm
that these and only these accessories are present.
From the PBS D
EDICATED NOZZLE PANEL you can click any of the following buttons:
Prepare: to acknowledge that all nozzle devices are present and that the desired
accessories are present to ensure a successful treatment. Click Prepare when
no red cross marks remain on this tab.
Note: Be aware that you must confirm the presence of the range shifter, ridge
filter, block, and range compensator if you want to proceed.
After clicking Prepare, you can still unconfirm the presence of the range shifter.
.
Unprepare: press this button if you wish to revert your previously taken decision
to acknowledge the nozzle condition.
Make sure that all required nozzle elements and accessories are present and click
Prepare. The colored state bar reads Ready (indicated by the green color).
WARNING Before allowing patient irradiation to start, check to see that the proper
irradiation accessory is installed on the nozzle
Using a different accessory than the one prescribed will result in
mistreatment of the patient.