Users Manual Part 4

Clinical User’s Guide
IBA | 35-11|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Preparing the Equipment for Irradiation
|
Verifying the PBS Dedicated Nozzle
The PBS DEDICATED NOZZLE PANEL of the EQUIPMENT PREPARATION SCREEN displays
two different tabs:
Overview: for information, refer to section “Overview” on page 35-11.
Nozzle Devices Details: for information, refer to section “Nozzle Devices
Details” on page 35-14.
Overview
On the OVERVIEW tab you have to ensure that all nozzle devices are present and
correctly set to ensure adequate treatment.
In addition, you have to ensure that all desired accessories, and only those
accessories, are present.
As long as the colored state bar reads Not Ready (indicated by the orange color), you
will not be able to proceed to irradiation.
When all positioning information is available to enable irradiation, the colored state
bar reads Ready (indicated by the green color).
Figure 35-7. PBS Dedicated Nozzle Panel
Overview tab
The O
VERVIEW tab contains the following feedback information:
For nozzle devices: