Users Manual Part 4
Clinical User’s Guide
IBA | 34-3|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Proceeding to Beam Selection
|
Figure 34-2. Beam Selection Screen
The header now displays the type of accessory, if any, and its identifier, as follows:
Pencil Beam Scanning:
No accessory: no accessory type
One accessory: accessory type and identifier
The list area is now filled with the beam(s) that you selected from the S
ESSION
DEFINITION SCREEN.
If there are multiple beams in the list, click the beam that you want to process first.
The selected beam appears in a highlighted box.
The associated data appear in the information area. You have to complete all
remaining treatment session steps with this beam and then you can select a next
beam from the list area.
Important In treatment mode, the approval to deliver a treatment beam to the
patient is under the responsibility of the Radiation Therapy
Technologist (RTT) as at Proton Therapy System (PTS) level no
distinction can be made between a calibrated/approved and a non-
calibrated/non-approved beam.