Users Manual Part 4
Clinical User’s Guide
IBA | 33-9|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Including One or More Beams into a Session
|
Distal Open Beam Range (g/cm
2
)
1
: the range of the proton beam farthest
away from the nozzle, without any beam modifying accessories (range
shifter, block, range compensator, etc.).
Accessories:
Block: identifier of the block (optionally used).
Compensator: identifier of the range compensator (optionally used).
Snout ID
Ridge Filter: identifier of the ridge filter (optionally used).
Range Shifter: identifier of the range shifter (optionally used).
Note: At least one accessory must be present.
Patient Positioning:
Patient Support Type: as specified in the beam; possible entries are:
Chair
Table
Patient Orientation: as specified in the beam; possible entries are:
Sitting
HFP: Head First-Prone
HFDR: Head First-Decubitus Right
FFDR: Feet First-Decubitus Right
FFP: Feet First-Prone
HFS: Head First-Supine
HFDL: Head First-Decubitus Left
FFDL: Feet First-Decubitus Left
FFS: Feet First-Supine
Gantry angle (deg): as obtained from the prescription.
Note: The gantry angle is always defined in the IEC 61217 fixed coordinate
system.