Users Manual Part 4

Clinical User’s Guide
IBA | 33-7|
Volume 1 - Treatment Session
|Part VI
- Using adaPT
deliver
in Standalone Mode Including One or More Beams into a Session
|
Reviewing Clinical Prescriptions - for PBS
From the Information area, select the Beam tab and the clinical data on that beam
appear. This is the clinical data that is prescribed in the treatment plan.
Figure 33-3. Beam Information - Pencil Beam Scanning
The clinical data is as follows:
Beam:
Name
Treatment Machine: the ID of the treatment room
Baselined by: the name of the user who baselined the beam.
Tolerance table
Delivery technique
Description
Geometry:
Proximal Open Beam Range (g/cm
2
)
1
: the range of the proton beam close
by the nozzle, without any beam modifying accessories (range shifter, block,
range compensator, etc.).
Maximum Field Size FX (cm): in PBS, extension of the beam along the X-
axis, specified by means of parameters FX, X1, X2, expressed in cm.