Users Manual Part 3

Clinical User’s Guide
IBA | 29-7|
Volume 1 - Treatment Session
|Part IV
- Using
ada
PT
prescribe
Managing Beams
|
Copy a beam from a Machine_QA plan and paste it into:
A plan of a QA-patient; this plan:
must be internal
must be non-baselined
must have the same patient orientation and patient support category
as the plan of origin
can have another TR
After pasting:
the beam is unbaselined
the beam can be edited
Copying a Beam
To copy a beam:
1. Right-click that beam. A pop-up menu appears.
2. Select Copy.
Pasting a Beam
To paste a beam:
1. Select the plan where you want to paste the beam.
2. Click Edit Patient to enter edit mode.
1. Right-click the plan. A pop-up menu appears.
2. Select Paste.
The message “The beam has been successfully copied” appears.
The copied beam appears with the name of the original beam, appended with
“_copy”. Optionally, you can rename this beam.
Saving a PBS Beam
When saving a PBS beam after editing and the requested amount of Monitor Units
(MU) per spot exceeds the configured clinical range for the maximum dose,
adaPTprescribe notifies you that a confirmation will be requested at the time of
baselining and treatment.
The clinical range is configured using adaPTprescribe in Administration mode (refer
to section “Managing MU Clinical Ranges” on page C-5).