Users Manual Part 3

Clinical User’s Guide
IBA | 28-7|
Volume 1 - Treatment Session
|Part IV
- Using
ada
PT
prescribe
Managing Plans
|
Figure 28-6. Couch Positions Check - Pop-Up Message
There is a configurable tolerance for the check (default value 2 mm), which can be
modified by IBA following a user request.
Click OK if you want to proceed with your selection, else click Cancel.
There is a user right: ISOCENTER_CHECK_OVERRIDE to authorize users to continue
with the workflow despite the consistency check finding an inconsistency.
For the configuration of the user rights in adaPTprescribe, refer to Appendix B
Managing PTS Users” on page B-1.
Confirming Baselining in PBS
Confirm the baseline action by entering your name and password.
Receiving the Baseline Notification
Once you have confirmed the baseline action by entering your name and password,
as it is the case with beam baselining, you receive a notification that the baseline
action has been performed successfully (see Figure 29-10).
You can only unbaseline a plan if none of the beams it contains has been irradiated.
To unbaseline a plan, click Unbaseline.
Copying and Pasting a Plan
Both a baselined or an unbaselined plan can be copied and pasted to another study
of the same patient or QA patient.
To be able to copy and paste a plan, click Edit Patient to enter editing mode.