Users Manual Part 2
Clinical User’s Guide
IBA | 14-21|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Installing and Removing Accessories into/from the Accessory Drawer or PBS
The first block slice has been installed into slot 1.1
The second block slice has been installed into slot 1.2
The third block slice has been installed into slot 1.3
The fourth block slice has been installed into slot 1.4
The second accessory has been installed into slot 2
Leftmost row:
The snout has been retracted
The snout is locked in place
Four switch detectors installed at the back of the snout detect the correct installation
of the blocks. A lock at the back of the snout (see Figure 14-8) keeps the slot 2
accessory (if any) in place and activates the snout lock.
Otherwise, the LEDs are OFF.
Figure 14-10. LEDs on the Snout
Important The gantry will not rotate unless an accessory is inserted in slot 1.4
(closest to the patient). Irradiation and snout holder motion will also be
prevented.
An empty frame (dummy aperture) is available in case the treatment
plan does not require any accessory in slot 1.4 of the snout.