Users Manual Part 2

IBA | 14-18 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Installing and Removing Accessories into/from the Accessory Drawer or PBS
Figure 14-7. Type DN_M Snout (lateral cut)
Removing Accessories from a DN_M Type Snout
The procedure to remove a block and accessory from a DN_M type snout is as
follows:
1. Pull the spring-loaded retaining pin to release the hinged retainer.
2. Open the hinged retainer.
3. Unlock the safety lock retaining the accessory installed on slot 2.
4. Remove the accessory from slot 2.
5. Unlock the safety lock retaining the block slice installed on slot 1.2.
6. Remove the block slice from slot 1.2.
7. Unlock the safety lock retaining the block slice installed on slot 1.1.
8. Remove the block slice from slot 1.1.
9. Push the surface of the hinged retainer to close it; the spring-loaded retaining
pin automatically falls into its locked position.
Safety locks
Hinged retainer
Spring loaded
retaining pin
presence
switches (3)
keys (2) (on
snout) &
keyways (on
accessory)
CAUTION Do not use the spring-loaded retaining pin to close the hinged retainer.
When you try to use the spring-loaded retaining pin, you may crush
your fingers.