Users Manual Part 2

Clinical User’s Guide
IBA | 14-17|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Installing and Removing Accessories into/from the Accessory Drawer or PBS
Installing Accessories onto a DN_M Type Snout
Note: The maximum combined weight of all accessories installed in the DN_M type
snout must not exceed 20 kg.
The procedure to install a block and accessory onto a DN_M type snout is as follows:
1. Pull the spring-loaded retaining pin to release the hinged retainer.
2. Open the hinged retainer.
3. Align the two key-ways on the edge of the first block slice with the two keys
inside the groove of the fixed block holder.
4. Insert the first block slice; the safety lock falls into place and the LED on the
snout’s tray indicating ‘Snout Slot 1.1’ turns green.
5. Repeat Step 3 with the second block slice.
6. Insert the second block slice; the safety lock falls into place and the LED on the
snout’s tray indicating ‘Snout Slot 1.2’ turns green.
7. Repeat Step 5 with the range compensator, range shifter or ridge filter to be
installed on the third slot.
8. Insert the range compensator, range shifter or ridge filter; the safety lock falls
into place and the LED on the snout’s tray indicating ‘Snout Slot 2’ turns green.
9. Push the surface of the hinged retainer to close it; the spring-loaded retaining
pin automatically falls into its locked position.
CAUTION Do not use the spring-loaded retaining pin to close the hinged retainer.
When you try to use the spring-loaded retaining pin, you may crush
your fingers.