Users Manual Part 2

IBA | 14-10 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Installing and Removing Accessories into/from the Accessory Drawer or PBS
The LEDs turn ON (green light) when the respective presence switches or pressure
pins send a signal indicating that (each item corresponding to a LED in the same
order as the list above):
The snout has been inserted
The block has been installed into slot 1
The second accessory has been installed into slot 2
The hinged retainer is closed
Two presence switches installed at the back of the snout detect the correct
installation of the accessories. A pressure pin on the hinged retainer detects whether
or not the retainer has been closed.
Otherwise, the LEDs are OFF.
Figure 14-3. LEDs on the Snout’s Tray