Users Manual Part 2
Clinical User’s Guide
IBA | 14-5|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Installing and Removing Accessories into/from the Accessory Drawer or PBS
Accessory Holder Operations
The accessory holder need not be in any particular position to enable an accessory
or the snout to be inserted or removed. On the other hand, for safety reasons,
moving the accessory drawer using the hand pendant is only possible when the
snout holder is retracted.
Note: With regard to ease of handling, the range shifter, snout type XL, and the
ridge filter can ideally be mounted onto and dismounted from the accessory drawer
with the gantry rotated to the 270° (9:00 o’clock) or 90° (3:00 o’clock) position.
Figure 14-1. PBS Dedicated Snout Holder
Installing an Accessory or Snout onto the Accessory
Drawer
The procedure to install an accessory or snout onto the drawer is as follows:
1. Make sure that the tray toggle switch is in the Unlocked position (see Figure 14-
1), i.e., up.
2. Insert the tray mounted accessory or snout into the drawer. Make sure that it is
completely introduced.
accessory drawer
tray (with
accessory)
tray lock
toggle switch
snout
holder
snout lock
toggle switch