Users Manual Part 2

Clinical User’s Guide
IBA | 13-21|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
Interrupting, Resuming, or Cancelling an Automatic
Docking or Undocking Operation
Interrupting an Automatic Docking or Undocking
Operation
To interrupt an ongoing automatic docking or undocking operation, release the
relevant Move button on the hand pendant.
Resuming an Automatic Docking or Undocking
Operation
To resume an interrupted operation:
1. Click OK to acknowledge the message that appears.
2. From the AUTOMATIC DOCKING SCREEN (see Figure 13-12), select the docking or
undocking function that was interrupted.
3. Touch and hold the relevant Move button till completion of the selected function.
4. Press OK to acknowledge the message that appears.
Canceling an Automatic Docking or Undocking
Operation
If you do not want a docking or undocking operation to complete orderly, perform as
follows:
1. Release the M (Move) button on the hand pendant to interrupt the ongoing
operation.
2. Click OK to acknowledge the message that appears.
3. Tap the Back button.
Note: The Back button is the rightmost button at the bottom of the hand
pendant, below the screen (see Figure 13-15).