Users Manual Part 2
IBA | 13-20 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
6. Disengage the docking device coupler fails to disengage from the PPS coupler:
a. Long press (approximately 3 seconds) the lock/unlock button on the wrist.
The button starts blinking.
b. Release the lock/unlock button on the wrist.
c. Within two seconds, press the lock/unlock button on the wrist until the
button goes dark (approximately 1 second).
d. Visually verify that the lock/unlock button on the wrist is dark.
Note: If the docking device coupler fails to disengage from the PPS
coupler, check to see that the lock/unlock button on the wrist is dark.
7. Press and hold the Motion Enable Button on the hand pendant.
8. Click the Release icon from the AUTOMATIC DOCKING SCREEN (see Figure 13-12).
9. Touch and hold the relevant Move button till completion of the Release
sequence.