Users Manual Part 2
Clinical User’s Guide
IBA | 13-17|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
10. Visually verify that the lock/unlock button on the wrist is lit.
11. Press and hold the Motion Enable Button on the hand pendant.
12. Click the Lift icon from the AUTOMATIC DOCKING SCREEN (see Figure 13-12).
13. Touch and hold the relevant Move button till completion of the Lift sequence.
Figure 13-12. Automatic Docking Screen
CAUTION Movement of the Patient Positioning System (PPS) is only possible if
the coupler is properly engaged and if the docking device is locked, i.e.,
the lock/unlock button on the coupling system has been pressed.
Lift
Fetch