Users Manual Part 2

IBA | 13-16 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
Figure 13-11. Advanced Menu
5. Press and hold the Motion Enable Button on the hand pendant.
6. Click the Fetch icon from the AUTOMATIC DOCKING SCREEN (see Figure 13-12).
7. Touch and hold the relevant Move button till completion of the Fetch sequence.
8. Click OK to acknowledge the message that appears.
9. Press the lock/unlock button on the wrist to manually lock the docking device
coupler and the PPS coupler.
Note: If the docking device coupler fails to engage with the PPS coupler, check
to see that the lock/unlock button on the wrist is lit.