Users Manual Part 2

Clinical User’s Guide
IBA | 13-15|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
Automatically Docking a Docking Device
To automatically attach a docking device to the PPS:
1. Move the trolley with the docking device mounted on it to the docking position.
2. Lock the trolley in that position.
3. Make sure that the lock/unlock button on the wrist is pressed.
Note: If the lock/unlock button on the wrist is not pressed at this stage, the
docking procedure will fail.
4. Select the Automatic Docking option from the ADVANCED MENU (see Figure 13-
11 ) . Th e A
UTOMATIC DOCKING SCREEN appears (see Figure 13-12).
CAUTION A total load on the docking device exceeding 180 kg (397 lbs) will
generate a prompt on the hand pendant requesting the Radiation
Therapy Technologist (RTT) for an acknowledgment to proceed.
Any load on the docking device exceeding 250 kg (551 lbs) will prevent
the PPS from moving.
CAUTION The Patient Positioning System (PPS) coupler and the docking device
coupler can only lock when they are properly aligned. Make sure that
both couplers are properly aligned before pressing the lock/unlock
button on the wrist.