- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
Cancel: to cancel the selected docking operation.
Note: Be aware that neither Manual nor Auto motions are possible before
termination of an automatic docking or undocking operation when the PPS is in
the docking area. If you cannot allow an ongoing docking or undocking operation
to terminate orderly, you must select the Cancel option if you want to perform
a Manual or Auto motion with the PPS.
Figure 13-10. Automatic Docking - Sequence of Operations
sideview
Loading Unloading
docking position docking position
Lift
Fetch
Put Back