Users Manual Part 2
IBA | 13-10 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
4. Keep the lock/unlock button on the coupling system pressed until it starts
blinking.
5. Release the button, then press it again. The docking device is now pneumatically
unlocked from the wrist.
6. Visually verify that the lock/unlock button on the wrist is not lit anymore.
7. Raise the docking device slowly over the coupler housing.
8. Long touch the PPS icon on the WIRELESS HAND-PENDANT GUI SCREEN (see
Figure 8-2). The P
OSITIONER MENU appears (see Figure 13-6).
9. Press and hold the Motion Enable Button on the hand-pendant.
10. Long touch Mode. Ready to Move appears.
11. Manually move the docking device away from the coupling system of the PPS.
CAUTION Make sure to sufficiently support the docking device before pressing
the lock/unlock button on the wrist to unlock the couplers. Failure to do
so may result in bodily harm or equipment damage.