Users Manual Part 2

Clinical User’s Guide
IBA | 13-9|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
7. Long touch the PPS icon on the WIRELESS HAND-PENDANT GUI SCREEN (see
Figure 8-2). The P
OSITIONER MENU appears (see Figure 13-6).
8. Press and hold the Motion Enable Button on the hand-pendant.
9. Long touch Mode. Ready to Move appears (Figure 13-7).
Figure 13-7. Positioner Menu
Ready to Move
Manually Undocking a PPS Support Device
To manually release a docking device from the PPS:
1. Move the PPS in a position that enables the docking device to be easily removed
from the treatment room.
2. Long touch the PPS icon on the WIRELESS HAND-PENDANT GUI SCREEN (see
Figure 8-2). The P
OSITIONER MENU appears (see Figure 13-6).
3. Long touch Mode. Ready to Undock appears.