Users Manual Part 2

IBA | 13-4 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Attaching/Removing a Patient Support Device to/from the PPS
|
Each docking device also has one air locking coupler installed, facing down. Joining
the docking device coupler with the corresponding PPS coupler firmly latches the
docking device to the PPS.
Figure 13-4. Lock/Unlock Button
Once the docking device is firmly latched to the PPS, pressurized air is required to
maintain the couch firmly locked with the robot and to ensure the repeatability of the
positioning of the couch on the PPS. However, the couch locking is failsafe. In case
of power failure, the couch will remain attached to the robot. It may, however, slightly
move with respect to the robot. In such a case, the accuracy of the patient
positioning cannot be guarantied, and no treatment shall be performed. Power must
be established to be able to disengage the couplers.
Lock/Unlock Button