Users Manual Part 2

IBA | 12-22 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Alignment Tools and Devices
|
Figure 12-4. CBCT Commands on the X-ray Console
Check the absence of collision before initiating the CBCT acquisition by performing
a dry-run. To perform the dry-run:
1. Select the CBCT acquisition parameter from the drop-down menu in
adaPTinsight (i.e., Clockwise or Counter ClockWise and define start angle). This
automatically defines the Prestart and Postend position on the hand-pendant.
2. Proceed as detailed in section “Preparing Equipment for a CBCT Acquisition” on
page 9-32.
3. During the dry-run gantry rotation, check that there is no collision and that the
clearance is sufficient with the patient and/or equipment, such as anesthesia or
respiratory equipment.
Command
buttons
LEDs