Users Manual Part 2

Clinical User’s Guide
IBA | 12-15|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Alignment Tools and Devices
|
The background becomes highlighted if the setting is not compatible with the
irradiation.
Note: The desired DID flat panel must be in place before attempting to acquire an
X-ray image.
4. The system is now ready for X-ray images to be taken.
CAUTION No Digital Imaging Device (DID) flat panel must be plugged or
unplugged from the associated power supply while the power supply
is powered on. If there is a need to connect or disconnect a DID panel,
the associated power supply must be switched off first. Else the DID
panel may suffer severe damage.
WARNING Over time, the quality of digital radiographs may become degraded,
suffer from an offset or a deterioration of the gain of the flat panels or
from defective pixels.
Periodic tests must be performed to ensure sufficient X-ray image
quality. The frequency shall take into account the foreseeable flat panel
deterioration by neutrons.
In case of recalibration and corrective maintenance, contact
immediately the assigned responsible team. For detailed information
about these procedures, refer to the Original Equipment Manufacturer
(OEM) documentation.