Users Manual Part 2
IBA | 11-40 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
Figure 11-45 shows the options present in the A
DVANCED SETTINGS MENU:
Figure 11-45. Advanced Settings Menu
(typical)
Reset motion controller: this function resets internal Positioning Management
System communications. You may need to reset the motion controller after a
particular error occurs. This is recommended in the relevant error messages.
Reset motion interlocks: this function enables you select all positioning
devices so that they are all reset in a single push of the SRCU Restart button.
This action is necessary after an emergency stop has occurred.
Reset factory positions: this function resets all positions to their factory
settings.
Note: For further details on how to recover the system after an emergency
stop refer to the Safety and Emergency Recommendations document.
2. Click on the desired option, as follows:
a. If you wish to reset the check manager, click the ‘Reset Check Manager’
option.
The Check manager has been reset message appears at the top of
the A
DVANCED SETTINGS MENU.
Alternatively,
a. If you wish to reset the SRCU, click the ‘Reset SRCU’ option.
The following pop-up message appears: