Users Manual Part 2

Clinical User’s Guide
IBA | 11-35|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
1. Click on the line corresponding to the angle to which you want to move the
gantry.
Alternatively, click any point of the circle of the gantry dial.
The angle corresponding to the line that you clicked or to the line that is closest
to the point of the circle that you clicked appears in the angle field in the center
of the gantry dial. The dot below the angle field becomes blue.
Figure 11-36. Gantry Commands - Target Angle Selected
(typical)
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the target angle.
To move the gantry clockwise, proceed as follows:
1. Click on the arrow pointing clockwise. The arrow becomes blue.
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the required position.
To move the gantry counterclockwise, proceed as follows:
1. Click on the arrow pointing counterclockwise. The arrow becomes blue.
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the required position.
Using Manual Mode to Move the PPS (Patient
Positioning System)
Note: Manual PPS motions are executed along the Horizontal Table Top Coordinate
System (HTTCS) axes.
The commands to move the PPS appear at the bottom of the Remote Positioning
Controls panel.