Users Manual Part 2

IBA | 11-34 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
Figure 11-34. Insert or Retract Snout Icon
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the required position.
Using Manual Mode to Move the Gantry
The commands to move the gantry appear in the middle of the Remote Positioning
Controls panel.
Figure 11-35. Gantry Commands
(typical)
These commands enable you to move the gantry in several ways.
To move the gantry to a particular angle, proceed as follows:
1. Click on the angle field in the center of the gantry dial, enter a target gantry angle
and press the Enter key on your keyboard or click the dot below the field. The
dot becomes blue.
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the target position.
To move the gantry to one of the angles available in the gantry widget, proceed as
follows:
Note: The gantry widget enables you to move the gantry to points every 1
between 0° and 345°.