Users Manual Part 2
IBA | 11-32 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
Figure 11-29. Accessory Drawer and Snout Commands
(typical)
Moving the Accessory Drawer
Note: The system only enables you to move the accessory drawer when the snout
is at a completely retracted position (the furthest possible from the isocenter).
To move the accessory drawer into the beam path, proceed as follows:
1. Click on the Move Accessory Drawer Into the Beam Path icon.
Figure 11-30. Move Accessory Drawer Into the Beam Path Icon
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the required position.
To move the accessory drawer out of the beam path, proceed as follows:
1. Click on the Move Accessory Drawer Out of the Beam Path icon.
Figure 11-31. Move Accessory Drawer Out of the Beam Path Icon
2. Press and hold the MEB and Move buttons on the remote positioning hardware
console until the device reaches the required position.