Users Manual Part 2

IBA | 18-2 |
Addendum
to the PTS User Documentation Set
|Part II
- Using Treatment Room Equipment Troubleshooting Treatment Room Equipment
|
Selected device DEVICE is in
unknown state. Probably it
has not been configured
properly.
The software controlling the device DEVICE has
not configured it properly.
Contact an operator to check its configuration.
Selected device DEVICE is
off!
The software controlling the device DEVICE has
performed a shutdown.
Contact an operator to restart it.
Selected device DEVICE is
initializing!
The software controlling the device DEVICE has
not configured it properly.
Contact an operator to check its configuration.
Latest corrected position has
not been reached. Press OK
to confirm
The latest computed correction vector has not
been completely applied to the PPS position.
Make sure the current PPS position is correct.
The positioner was
overloaded and may require
maintenance. Contact IBA
Support.
The load on the loadcell exceeded the maximum
allowed by the manufacturer specification. This
may have damaged the PPS. Contact an operator.
Emergency stop button
pressed.
An emergency button has been pressed.
Locate the pressed button and release it if the
emergency situation is over.
Perform a ‘clear emergency’ sequence to be able
to move devices again.
Lock accessory before any
gantry motion.
The accessory installed on the snout holder is not
properly locked. Lock it to be able to move the
gantry.
Lock accessory before any
drawer motion.
The accessory installed on the snout holder is not
properly locked. Lock it to be able to move the
drawer.
Lock accessory before any
snout motion.
The accessory installed on the snout holder is not
properly locked. Lock it to be able to move the
snout.
DEVICE is in maintenance
mode.
Device DEVICE is in maintenance mode. You
have to set it back to default mode to be able to
move it again.
DEVICE is in Error. Contact
operator.
An internal error occurred on device DEVICE.
Contact an operator.
Positioner motion is forbidden
due to docking mode. Please
disable this mode before
motion.
The PPS is still in docking mode.
Switch to move mode.
Message Description