Users Manual Part 2
Addendum
IBA | 18-1 |
to the PTS User Documentation Set
Chapter 18
Troubleshooting Treatment Room
• • • • • •
Equipment
The following table describes the error/warning messages displayed on the wireless
hand-pendant screen.
.
Message Description
DEVICE
a
must be homed
first.
Home the device DEVICE before performing this
action.
Gantry position too far from
prescription.
Position cannot be confirmed.
Move the gantry closer to the treatment position
before confirming the current position.
Lock the couch before
moving.
The couch is not properly locked. Lock it before
moving the PPS.
Positioner must be tared first. Remove everything from the couch and perform
a taring procedure in order to be able to move the
PPS.
Before taring, add a couch. There is no couch installed on the PPS. You have
to install a couch before taring.
Before taring, add a known
couch.
The couch installed on the PPS is not a valid
couch to tare. You have to install a valid couch
before taring. The validity of the couch is
assessed by the couchId value.
Selected device DEVICE is
unhealthy!
The device DEVICE has an internal error
preventing it from moving. Contact an operator.
Selected device DEVICE is
faulty!
The device DEVICE is outside valid motion range.
You have to perform a recover sequence before