Users Manual Part 2
IBA | 11-22 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
Figure 11-17. CBCT Preparation Panel - Equipment Ready
(typical)
4. You may now continue the CBCT acquisition procedure using adaPTinsight.
Note: For further details on CBCT acquisition, refer to the adaPTinsight user
documentation listed in the Delivery Note.
Using Auto Mode to Move to a User Preset
In order to move the patient positioning devices to a user preset, proceed as follows:
1. Select the required preset from the ‘User Presets’ part of the Remote
Positioning Controls panel.