Users Manual Part 2

Clinical User’s Guide
IBA | 15-1 |
Volume 1 - Treatment Session
Chapter 15
Resetting a Dose Counter Electronic
• • • • • •
Unit
The Dose Counter Electronic Unit (DCEU) is a redundant dose counter. It is used to:
Provide a redundant reading of the dose monitor units received by the patient.
Provide a redundant way of stopping the irradiation when the target dose
entered on the Therapy Control System (TCS) is reached.
Record the counter value in case of computer failure or power failure, thus
allowing the therapist to note the dose delivered at the time of failure.