Users Manual Part 1

IBA | 2-4 |
Clinical User’s Guide
Volume 1 - Treatment Session
Starting Up Applications in the adaPT
Treatment Suite
|
Characteristics of the Clinical Environment
Each of the applications in the adaPTsuite, i.e., adaPTprescribe, adaPTinsight, and
adaPTdeliver, requires you to perform a dedicated login.
When the applications in the adaPTsuite have been validated for clinical use, the
background color of the header is black. This is indicative of a Clinical environment.
Whenever the applications in the adaPTsuite have not been validated for clinical use,
the background color of the header is red. This is indicative of a non-Clinical
environment. In these circumstances, the applications are NOT for clinical use. The
applications must not be used for treating patients.
Note: Within the Clinical environment, the different adaPTsuite applications offer
different log-in modes. AdaPTdeliver in particular offers a Clinical and a Physics mode.
AdaPTdeliver's Physics mode is to be used for QA purposes, calibration and research
activities only.
WARNING Any formal activity (e.g., treatment, patient Quality Assurance (QA),
clinical acceptance, etc.) performed out of the Clinical environment is
forbidden. The Proton Therapy System cannot be guaranteed to satisfy
safety and performance criteria needed for such activities out of the
Clinical environment.