Clinical User’s Guide
IBA | 2-3|
Volume 1 - Treatment Session
Starting Up Applications in the adaPT
Treatment Suite
|
Each morning, and before beginning clinically related operations, it is recommended
to verify that the content of the ‘About TCS’ icon stored in the MCR log book by IBA
operators is not unexpectedly modified compared to the previous day of clinical use.
Figure 2-1. ‘About TCS’ Icon Content
The Clinical Environment
Global
Checksum
WARNING In case of any unexpected change of the Global Checksum line
compared to the line present in the ‘About TCS’ content of the previous
day of clinical use, you must suspend all clinically related activities. The
safety and essential performances of the Proton Therapy System
cannot be guaranteed. In this case, you must contact IBA immediately
for support.
WARNING The Clinical environment is the only suitable environment for clinically
related activities.