Users Manual Part 1

Clinical User’s Guide
IBA | 2-1 |
Volume 1 - Treatment Session
Chapter 2
Starting Up Applications in the adaPT
• • • • • •
Treatment Suite
Startup Operations
At startup in the Main Control Room (MCR) and TCR, the PTS performs automated
checks on the versions of critical software running in Control Units (CU) and
Electronic Units (EU) inside the system against the expected versions of these
software packages.
In case of a mismatch a pop-up message appears and the PTS blocks startup.
Clinical and Non-clinical Environment
WARNING In case you opt to bypass the software version checks, i.e., you click OK
to acknowledge that you want to continue, knowing that such version
discrepancies exist, be aware that any formal activity (e.g., treatment,
patient Quality Assurance (QA), clinical acceptance, etc.) is forbidden.
WARNING Any formal activity (e.g., treatment, patient Quality Assurance (QA),
clinical acceptance, etc.) performed out of the Clinical environment is
forbidden. The Proton Therapy System cannot be guaranteed to satisfy
safety and performance criteria needed for such activities out of the
Clinical environment.