Users Manual Part 1

IBA | 1-14 |
Clinical User’s Guide
Volume 1 - Treatment Session
Introduction
|
For detailed information on the use of adaPTdeliver that is generic to both Clinical
mode and Physics mode, refer to Part V, “Using adaPTdeliver.
For detailed information on the use of adaPTdeliver specifically in Physics mode,
refer to Part IX, “adaPTdeliver in Physics Mode.
Using Information From the OIS
You can select a patient, study, plan, or beam using an OIS connection but the OIS is
not informed about any such retrieval, nor does the PTS send any Physics mode
related information to the OIS.
When the OIS sends a plan in Physics mode, the PTS issues an error message to the
OIS that the plan is erroneous. In this way, the OIS is prevented from storing any
subsequent data. The plan, however, is used in PTS.
The steps to transfer a plan from the OIS to the PTS are identical in Physics mode
and Clinical mode. No special or extra actions are required.