Users Manual Part 1

IBA | 11-8 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
Figure 11-4. Error Tool Tip
(example)
Operating Modes: Auto and Manual
The remote positioning controls can be used, by default, in a single operating mode
in adaPTdeliver clinical sessions:
Auto mode: using Auto mode, you may perform Auto motions to prescribed or
user-defined target positions.
For detailed information on Auto motions, refer to Chapter 6.
Note: Only those target positions for which conditions are currently met are
available using Auto mode.
Auto mode also gives you access to the motion of the imaging devices.
The Manual operating mode may be available in the remote positioning controls of
adaPTdeliver clinical sessions at your site upon explicit request to IBA:
Manual mode: using Manual mode, you may perform Manual motions to
positions that are not predefined.
For detailed information on Manual motions, refer to Chapter 6.
Note: For the purpose of this manual, the Manual operating mode of the remote
positioning controls is fully explained and illustrated. This operating mode may or
may not be available at your site.
Both Manual and Auto operating modes are available in adaPTdeliver physics mode
sessions.
Manual and Auto operating modes are available by default in the wireless hand-
pendant GUI. For more details on how to use the wireless hand-pendant, refer to
Chapter 9.
Using the Remote Positioning Controls in
Auto Mode
Auto mode enables you to perform Auto motions to patient-specific and non patient-
specific target positions.