IBA | 11-7|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Remote Positioning Controls
|
If an emergency stop button local to your treatment room is pressed during Patient
Positioning Device (PPD) movement, the motion stops immediately.
After using an emergency stop button local to your treatment room to stop the PPDs,
the emergency stop button and the Therapy Safety System (TSS) must be reset.
After the emergency stop button and TSS are reset, the following may also be
required:
Recovery enabling beam production
Recovery of gantry movement (for GTR only)
Recovery of PBS Dedicated snout and PPS movements
Recovery enabling beam in TR
Recovery enabling X-ray high-voltage
Verification of proper operation
Successful daily calibrations (e.g., variable collimators)
Using Tool Tips
The remote positioning controls panel consists of tool tips.
Place the mouse pointer over any given icon to see a tool tip with some details
regarding what that particular icon or button can enable you to do, the meaning of its
current state, etc.
Figure 11-3. Tool Tip
(example)
Tool tips also exist for error messages.
Place the mouse pointer over an error sign to see a tool tip with some details
regarding the occurring error.