Users Manual Part 1

IBA | 9-38 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
Figure 9-30. CBCT Dry Run Screen
(typical)
5. Use the Move buttons on the CBCT DRY RUN SCREEN to perform a CBCT dry run
before proceeding to CBCT acquisition using adaPTinsight.
Note: You may perform the following actions in the order you consider
appropriate considering the current circumstances. The order explained here is
a suggestion.
a. Use the Move Gantry to Postend Position button to move the gantry to
the position reached at the end of a CBCT acquisition.
b. Use the Move Gantry to Prestart Position button to move the gantry to
the position required to start a CBCT acquisition.
The gantry is now in position to proceed to CBCT acquisition using adaPTinsight.
Note: For further details on CBCT acquisition, refer to the adaPTinsight user
documentation listed in the Delivery Note.
Move Gantry to
Prestart
Position button
Move Gantry to
Postend
Position button