Users Manual Part 1

Clinical User’s Guide
IBA | 9-31|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
3. While holding the PPS + Pitch or PPS - Pitch button, press and hold the MEB
buttons on the wireless hand-pendant until the device reaches the required
position.
Rolling the PPS
To roll the PPS, proceed as follows:
1. To access the PPS pitch and roll commands, touch the widget located above the
central dial, on the left of the screen.
2. Touch and hold the PPS + Roll button to roll the PPS in the positive direction.
Alternatively, touch and hold the PPS - Roll button to roll the PPS in the negative
direction.
Figure 9-22. PPS + Roll Icon (above) and PPS - Roll Icon (below)
Note: The arrows indicate the rotation around the roll axis.
3. While holding the PPS + Roll or PPS - Roll button, press and hold the MEB
buttons on the wireless hand-pendant until the device reaches the required
position.
Moving the Digital Imaging Devices
Digital imaging devices options may be accessed either from the Auto or Manual
mode screen. The options available in both modes are identical.
1. Touch the Digital Imaging Devices icon on the device selection bar at the
bottom of the screen.
Once the digital imaging devices subsystem is selected, the interface displays
four Move buttons in the central dial:
The top button enables you to insert the devices needed for performing
stereoscopic image acquisition.
The left button enables you to insert the devices needed for performing
orthogonal image acquisition.
roll axis