Users Manual Part 1

Clinical User’s Guide
IBA | 9-29|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
Moving the PPS along the Y Axis
To move the PPS along the Y axis, proceed as follows:
1. To move the PPS in the + direction along the Y axis, touch and hold the Y+
button.
Alternatively, to move the PPS in the - direction along the Y axis, touch and hold
the Y- button.
2. While holding the Y+ or Y- button, press and hold the MEB buttons on the
wireless hand-pendant until the device reaches the required position.
Moving the PPS along the Z Axis
To move the PPS along the Z axis, proceed as follows:
1. Touch and hold the Z tab on the PPS Z slider and slide it in the desired direction
(positive movement along the Z axis if the slider is moved clockwise, negative
movement along the Z axis if the slider is moved counterclockwise).
2. While holding the Z tab, press and hold the MEB buttons on the wireless hand-
pendant until the device reaches the required position.
Rotating the PPS Around the Isocenter
To rotate the PPS around the isocenter, proceed as follows:
1. Touch and hold the ROT tab on the PPS ROT slider and slide it in the desired
direction (clockwise PPS rotation when looking at the couch from above if the
slider is moved to the left, counterclockwise PPS rotation when looking at the
couch from above if the slider is moved to the right).
2. While holding the ROT tab, press and hold the MEB buttons on the wireless
hand-pendant until the device reaches the required position.
Pitching the PPS
To pitch the PPS, proceed as follows:
1. To access the PPS pitch and roll commands, touch the widget located above the
central dial, on the left of the screen.