Users Manual Part 1
IBA | 9-26 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
2. Hold the ROT tab until the device reaches the required position.
3. While holding the ROT tab, press and hold the MEB buttons on the wireless
hand-pendant until the device reaches required position.
Moving the Accessory Drawer
Note: The system only enables you to move the accessory drawer when the snout
is in a completely retracted position (the furthest possible from the isocenter).
To move the accessory drawer into the beam path, proceed as follows:
1. Touch and hold the Move Accessory Drawer Into the Beam Path button.
Figure 9-15. Move Accessory Drawer Into the Beam Path Icon
2. While holding the Move Accessory Drawer Into the Beam Path button, press
and hold the MEB buttons on the wireless hand-pendant until the device reaches
the required position.
To move the accessory drawer out of the beam path, proceed as follows:
1. Touch and hold the Move Accessory Drawer Out of the Beam Path button.
Figure 9-16. Move Accessory Drawer Out of the Beam Path Icon
2. While holding the Move Accessory Drawer Out of the Beam Path button,
press and hold the MEB buttons on the wireless hand-pendant until the device
reaches the required position.
Moving the Snout holder
To insert the snout, proceed as follows:
1. Touch and hold the Insert Snout button.