Users Manual Part 1

IBA | xviii |
Clinical User’s Guide
Volume 1 - Treatment Session
Wireless Hand-Pendant
Generic:
Safe and effective use of the wireless device:
The wireless hand-pendant subsystem contains:
A Bluetooth radio with the following characteristics:
Bluetooth Low Energy (BLE) version 4.1 operating in the 2.400 GHz-
2.4835 GHz ISM band using F.H.S.S. modulation.
Effective Isotropic Radiated Power (EIRP): 0 dBm 1 mW Class 1.
A Smartphone using an IEEE 802.11 Wi-Fi link. Other wireless links in
the smartphone are factory disabled to avoid interferences.
Wi-Fi operating in the 2.400 GHz-2.4835 GHz ISM band using F.H.S.S.
modulation; maximum power 100 mW (20 dBm).
Wireless Quality of Service (QoS): the subsystem has been designed to be
fail safe (to stop movements) in case of loss of communication or significant
degradation. Only IBA authorized devices can connect to the hand-pendant
wireless access points in treatment room.
The battery of the wireless hand-pendant must be replaced by authorized
IBA personnel only.
Wireless security measure: The subsystem will only accept a connection
from a single IBA Hand-Pendant at a time in a given room with IBA
Universally Unique Identification protocol. The UUID is factory set and can
not be changed by the user. The IEEE 802.11 Wi-Fi link uses WPA2 for
authentication. Cross talk between devices in the same vicinity: Due to the
unique ID mechanism, it is not possible to operate two devices in the same
treatment room. Due to treatment room neutron shielding and low power
Bluetooth usage, crosstalk between equipments from outside treatment
rooms is very unlikely.
However we recommend to keep the wireless hand-pendant in the docking
station when not in use.
USA:
FCC PART 15 Information to user.
FCC ID: 2AHZSHP-MOB
FCC ID: 2AHZSHPV3C-MOB
FCC ID: 2AHZSHP-FIX