- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
Using the GUI in Auto Mode
Auto mode enables you to perform Auto motions to patient-specific and non patient-
specific target positions.
To select Auto mode, proceed as follows:
1. Touch the Auto option on the hand-pendant mode selection bar.
Auto option becomes highlighted by a blue line and the default Auto mode
screen appears.
Auto Mode Target Positions
A target position may be one of the following:
User preset: a user preset is a target position defined by the user. The label of
a user preset is the name of the preset as defined by the user.
Setup position: setup positions are defined in the Treatment Plan and
associated to setup beams. As such, a setup position is only available as a target
position when a setup beam is selected and the preparation for the setup beam
has started.
Treatment position: treatment positions are associated to treatment beams.
As such, a treatment position is only available as a target position when a
treatment beam is selected and the preparation for the treatment beam has
started.
Corrected position: a corrected position becomes available once a given
treatment or setup position has been corrected using the Patient Position
Verification System (PPVS), adaPTinsight. As such, a corrected position is only
available as a labeled target position when a treatment or setup beam is selected
and after the alignment process is complete.
Depending on the stage of the workflow, you may choose among the available target
positions.
Selecting a Target Position
All prescribed and user-defined target positions are available in the menu accessible
via the Menu icon located at the top left of the screen.