Users Manual Part 1
Clinical User’s Guide
IBA | 9-5|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
controls that match the current room configuration. For this reason, there are as
many possible GUI configurations as there are room configurations. The same
wireless hand-pendant can be used in any treatment room as long as it has been
successfully registered to that treatment room.
The registration process guarantees that only one wireless hand-pendant is allowed
to send commands in a given treatment room.
The registration of a wireless hand-pendant to a given room is not possible if another
user is pressing the motion enable buttons on an already registered wireless hand-
pendant, irradiation is ongoing or the registration of another wireless hand-pendant
is ongoing. This rule guarantees that it is not possible to interfere with a user sending
commands to the Proton Therapy System.
When you enter a Treatment Room with a wireless hand-pendant that is not
registered, the wireless hand-pendant automatically prompts you to register it by
displaying the Register button, which enables you to start the registration procedure.
Although several wireless hand-pendants may be present in a given treatment room
and they may all be connected to the wireless network, only one wireless hand-
pendant can be registered to a single treatment room at any given time. The
registered wireless hand-pendant is the only one authorized and able to move
treatment room equipment.
Registration Procedure
In order to register a wireless hand-pendant to a given treatment room, proceed as
follows:
1. Once the wireless hand-pendant is in the treatment room, it automatically
connects to the treatment room Wi-Fi network. The following screen appears: