Users Manual Part 1
IBA | 9-4 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
Registering the Wireless Hand-Pendant
In order to use the wireless hand-pendant in a given treatment room, it is necessary
to register it to that treatment room.
When as part of the registration process the application connects via Wi-Fi to the
Proton Therapy System, it retrieves the room configuration of the room to which it is
currently registered. The room configuration includes the list of the installed devices
and their type. Based on this information, the application adapts itself to provide
Important When leaving the treatment room, you must always put the wireless
hand-pendant back on its docking station. This ensures that the
wireless hand-pendant remains protected from irradiation during
treatment.
Important The wireless hand-pendant is a failsafe device. In case of interruption
of the communication between the wireless hand-pendant and the rest
of the Proton Therapy System, any ongoing movement will be stopped.
The loss of communication of the wireless hand-pendant is considered
as a release of any currently pressed or tapped button. More
specifically, if the system looses the communication with the wireless
Motion Enable Button during a movement, it will stop and disable all
motion in the treatment room.
WARNING Use the wireless hand pendant always with the delivered IBA necklace
or with an alternative reliable means to prevent the device falling down
accidentally
Falling of the wireless hand pendant can cause severe damages,
including
- Damage to the battery
- Degradation inducing hazardous faulty behaviour.