Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Using the Wireless Hand-Pendant
|
Safety Measures Specific to the Use of the
Wireless Hand-Pendant
Make sure you read and understand the following warning and important messages
before using the wireless hand-pendant.
WARNING It is not allowed to activate movements with the wireless hand pendant
when there is no visibility on the moving devices (e.g., it is not allowed
to move the patient positioning devices from the maze).
WARNING Stop using the wireless hand pendant and inform IBA personnel in case
any unusual defect is observed on the wireless hand pendant.
Defects can be (but are not limited to) unusual overheating of the
device, leakage of substances, emission of fumes or gas, cracks in the
hand pendant case.
Important Other wireless devices could interfere with the wireless hand-pendant
and are therefore proscribed in the treatment room. Interferences could
lead to untimely and unexpected motion interruptions.
While remaining safe, such situation could impede the normal use of
the Proteus 235 system.
If you experience loss of effectiveness or an unexpected stop of
movements with the wireless hand-pendant, please move away
emitters like mobile phones, laptops, etc.
Important Do not use the wireless hand-pendant for another purpose than
controlling Proteus 235 system. Do not connect the wireless hand-
pendant manually to any Wi-Fi network. Do not connect the wireless
hand-pendant to another Wi-Fi network than a Proteus PLUS treatment
room.