Users Manual Part 1
Clinical User’s Guide
IBA | 8-13|
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Controlling Movement: Wireless Hand-Pendant and Remote Positioning Controls
|
the Locked - TCR hand-pendant in use. message appears on the
wireless hand-pendant display
Note: It is impossible to take over command when the Motion Enable Button
(MEB) is pressed on the active hand-pendant.
Note: It is impossible to take over command when the Move button on the X-ray
generator hardware console is being pressed. In cases where the Cone-Beam
Computed Tomography (CBCT) controls on the X-ray generator hardware console are
locking the control, press the Reset button on the console.
Switching Command From the Remote
Positioning Controls to the Wireless Hand-
Pendant
When you are using the remote positioning controls, the Locked - TCR hand-
pendant in use message appears on the wireless hand-pendant display.
To switch command to the wireless hand-pendant, squeeze and release the Motion
Enable Button (MEB) on the sides of the wireless hand-pendant:
the wireless hand-pendant is enabled
the remote positioning controls are disabled
the Locked - TR hand-pendant in use. message appears on the remote
positioning controls.
Note: It is impossible to take over command when the Motion Enable Button
(MEB) is being pressed on the remote positioning hardware console.