Users Manual Part 1

IBA | 7-16 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Moving the Patient Positioning Devices
|
d. Press OK
e. Move the equipment causing the collision. During this movement, collision
detection is disabled and equipement movements can be performed at low
speed only.
f. If this movement brings the system into a state in which the collision is no
longer present, see Step 2.
g. If you interrupt the movement while the collision is still present, see Step 3.
Managing Proximity
When a proximity situation is detected:
Patient Positioning Devices (PPDs) move at medium speed.
The wireless hand-pendant or the remote positioning controls make(s) a beeping
sound during equipment motion to alert you that the devices are getting close
to each other.
Note: If the PPDs are close but the motion is such that one is getting away of the
other, there is no proximity situation.