Users Manual Part 1
IBA | 7-12 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Moving the Patient Positioning Devices
|
CBCT menu -
Safe
Isocentric PPS Defines a target position for
the PPS. This position is only
available when the PPS is not
in a safe position for CBCT; this
position is used when the
current PPS position will lead
to a collision trajectory during a
CBCT scan. Moving the PPS
after choosing the CBCT Safe
position will move the PPS to a
position where no collision
occurs with the table.
Non-patient specific
LOAD FRS,
Isocentric
Gantry
PPS
PBS ded.
snout
Position used to load/unload a
patient onto/from the couch.
USER FRS Gantry
PPS
PBS ded.
snout
This position is only available if
in Service session and
previously defined and saved
in the TR when the PPD
settings were saved during a
“Save Position” request.
GANTRY 90 FRS,
Isocentric
Gantry Position used to move the
gantry to a frequently required
position.
GANTRY 180 FRS,
Isocentric
Gantry Position used to move the
gantry to a frequently required
position.
GANTRY 270 FRS,
Isocentric
Gantry Position used to move the
gantry to a frequently required
position.
Table 7-1. Labeled Positions for a GTR (Cont’d)
Position name Coordinate
system in
which it is
expressed
PPD Settings for target position