Users Manual Part 1
IBA | 7-2 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part II
- Using Treatment Room Equipment Moving the Patient Positioning Devices
|
WARNING The Leoni Orion maintenance hand pendant is a service hand pendant
only. As a service hand pendant, the Leoni Orion maintenance hand
pendant is capable of overriding safety checks.
It is strictly forbidden to use the Leoni Orion maintenance hand
pendant after an initial beam request has been completed, and for any
clinical operation in general.
Do NOT use the Leoni Orion maintenance hand pendant while the
patient is loaded.
WARNING It is strictly forbidden to ever place any objects (tools, parts, etc) on top
or inside moving devices such as the nozzle, or the gantry. Falling tools
or parts can injure patients or staff.
Take inventory of the tools or parts after using them in a treatment
room.
CAUTION In order to avoid Patient Positioning System (PPS) damage, do not use
the PPS robot as a lifting tool.
CAUTION Take caution for flat panel arm and gantry movement in GTR as they do
not respect the rotation and linear movement stopping distances.